In our daily work, how should we calibrate the flat panel detector

In our daily use, we will regularly calibrate the

flat panel detector

. Is it possible to calibrate the flat panel detector at any time of the day? The answer is: No.

Because the flat panel detector maintains the best working condition near the calibration temperature, the system also requires the

flat panel detector

temperature to be kept within a certain range of the calibration temperature. If the temperature exceeds the range, there may be an error or even no exposure. With continuous work, the temperature of the detector will rise to a certain extent.flat panel detectors

Therefore, it is best not to calibrate when it is just turned on or when it has not been used for a long time. It is also best not to do it immediately after a long and high-load work. It should be after the flat panel detector has worked for a period of time and waited for about 10 minutes before proceeding. This can ensure that the calibration temperature is at the interruption of the working temperature, and it is not easy to cause the temperature of the flat panel detector to exceed the required range of the calibration temperature.